CEL-400 Series Real Time Analyzers
|
Introduction
|
|
Key benefits |
|
|
The new CEL-400 series is an advanced and comprehensive
instrument family for a wide range of applications. It has been designed for
convenience and ease of use in many popular tasks. From simple workplace
noise surveys to full community noise audits the new CEL-400 can do it
quickly and easily. Advanced digital signal processing technology enable the
meters in the range to cover the whole audible sound range in a single span.
This eliminates the need to select the best range for a measurement since
everything is covered from the quietest quiet to the loudest loud. |
|
q Single ultra wide 140 dB dynamic
range q Real time octaves q Real time third octaves q 3 Simultaneous broadband frequency
weightings q Available in both ANSI Type 1
& Type 2 accuracy classifications q User definable setup menus for the
most commonly used measurements q Large and easy to read display q Intuitive keypad operation q Huge memory stores all that is
usually needed for most measurement requirements q Rapid data capture up to 100
spectra per second |
|
Applications
|
|
|
|
|
Workplace noise: Automatic calculation of noise exposure. Real time octave
band measurements for the correct selection of hearing protectors. Community noise: All the necessary parameters measured for site boundary
noise, impact assessments, and transportation noise studies. Available with a
wide range of accessories for unattended monitoring. Product noise: New product development, emission assessment, quality
control, and noise reduction troubleshooting. General noise: All types of steady and variable or impulsive noise can be
quickly and accurately assessed using the real time frequency filters in the
B and C models of the meters. |
The primary role of the CEL-450
model is an occupational hygiene instrument. It can be purchased as a
broadband model that automatically calculates noise exposure (TWA) and can be
used for workplace noise assessments. For the selection of the correct
hearing protection the CEL-450
features the calculation of LCeq-LAeq for use in the HML or NRR method. A
more detailed assessment of the frequencies of noise an employee is exposed
to can be made with the octave band model. For product development the 1/3
octave version will measure the more ‘tonal’ components of noise,
giving better assessment of noise emissions and allowing noise reduction
techniques to be applied more effectively. In frequency analysis modes the
instrument measures all the bands simultaneously in real-time, making for
fast and accurate results. |
The CEL-450 will
also measure time history profiles of noise that show how the noise climate
has changed over time. The stored results can then be downloaded, and
displayed on the included dB23 software. The main role of the CEL-490 is as an environmental instrument
for longer-term community noise applications. This model adds the LN%
statistical measurement parameters required for many environmental
regulations and these can be calculated in real time in both broadband and
frequency bands on the relevant versions. The CEL-490
also adds more comprehensive data logging functions in the form of period
time intervals selectable from 10 msec to 1 hour as well as time history
profile recording. Automatic start and stop timers are standard functions on
the CEL-490 and with the companion
outdoor microphone enclosure system it can be used for unattended monitoring. |
|
Product Information
|
CEL-450 |
CEL-490 |
|
|
Broadband
measured parameters |
L, Leq, Lmax, Lmin, Lpk,
Ltm3, Ltm5, LEP,d, TWA, LAE, Lavg,
LAeq-LCeq |
L, Leq, Lmax, Lmin, Lpk,
Ltm3, Ltm5, LEP,d, TWA, LAE, Lavg,
LAeq-LCeq, 5 x LN% (user selectable 0.1 to
99.9%) |
|
|
Octave and
third octave measured parameters |
L, Leq, Lmax, Lmin, Lpk |
L, Leq, Lmax, Lmin, Lpk,
5 x LN% (user selectable 0.1 to 99.9%) |
|
|
Timer
facility |
Fixed duration timer (1 min to 24 hours) |
Fixed duration timer (1 min to 24 hrs) Automatic On/Off timers (7 sets, up to 1 month in advance) |
|
|
Profile
time history |
4 broadband parameters measured (10 msec to 30 min) |
4 broadband parameters measured (10 msec to 30 min) |
|
|
Period
time history |
Not available |
All parameters selectable in broadband, octave and third
octave depending on model (10 ms to 1 hr) |
|
Specification
|
|
||
|
Acoustic accuracy: |
ANSI S1.4 (R1997), IEC 61672: 2002, IEC 60651:1994, IEC
60804: 2000, IEC 61260:Class 0 |
||
|
Frequency weightings: |
A, C and Z (unweighted) |
||
|
Time weightings: |
Slow, Fast and Impulse plus Peak |
||
|
Amplitude weightings: |
Exchange rate Q = 3 plus one from 4, 5 or 6 or none |
||
|
Measurement range: |
0 to 140 dB rms. Or 143.3 dB peak |
||
|
Noise floor: |
17 dBA – Class 1 models 25 dBA – Class 2 models |
||
|
Frequency analysis mode: |
Octave – 11 bands from 16 Hz to 16 kHz plus 3
broadbands A, C and Z Third octave – 33 bands from 12.5 Hz to 20 kHz plus
3 broadbands A, C & Z |
||
|
Stored user setup configurations: |
4 per mode with user selectable name plus factory default
(automatically used next time meter is used) |
||
|
Memory: |
2 M byte capacity stores up to 999 separate runs plus 880,000 broadband results, 40,000 octave spectra, 13,300
spectra |
||
|
Physical properties: |
Weight – 17 oz (480 gm) Size – 13.5 x 4 x 1.5 in (343 x 100 x 37 mm) Batteries – 4 x AA cells (alkaline or NiMH types
recommended) External power - 12 V dc at 150 mA from mains transformer
or battery pack Tripod socket – standard camera tripod thread
¼” Whitworth |
||
|
Pc software: |
Windows™ compatible package dB23 and RS232 cable
included with every model for download of stored results to pc and upload of
control to meter |
||
Ordering information
|
|
||
|
ANSI/IEC model compliance: Real time analyzer Type 1: Real time analyzer Type 2: |
Model part number: CEL-450.A1 CEL-450.B1 CEL-450.C1 CEL-490.A1 CEL-490.B1 CEL-490.C1 CEL-450.A2 CEL-450.B2 CEL-450.C2 CEL-490.A2 CEL-490.B2 CEL-490.C2 |
Functions available: Broadband only Broadband plus RT Octaves Broadband plus RT Octaves & 1/3 Octaves Broadband only Broadband plus RT Octaves Broadband plus RT Octaves & 1/3 Octaves Broadband only Broadband plus RT Octaves Broadband plus RT Octaves & 1/3 Octaves Broadband only Broadband plus RT Octaves Broadband plus RT Octaves & 1/3 Octaves |
|
|
Kits and other accessories: Standard kits are available including the relevant
real time analyzer, a matching CEL-110 acoustic calibrator, a foam
windscreen, RS232 cable, dB23 Windows™ software in a custom
attaché case by specifying the part number CEL-4X0.XX/K1 where CEL-4X0.XX
is the version of the instrument. Other accessories such as microphone extension cables,
long-term waterproof case; interface connections and power supply options are
available upon request. |
|||